External quality assessment (EQA) for CoaguChek monitors

Journal:Thrombosis and Haemostasis
ISSN:0340-6245
DOI:http://dx.doi.org/10.1160/TH09-10-0683
Issue:2010: 103/5 (May) pp. 875–1108
Pages:936-941

External quality assessment (EQA) for CoaguChek monitors

J. Jespersen (1), L. Poller (2), A. M. H. P. van den Besselaar (3), F. J. M. van der Meer (4), G. Palareti (5), A. Tripodi (6), C. Shiach (7), M. Keown (2), S. Ibrahim (2)

(1) Department of Clinical Biochemistry, Hospital of South West Denmark, Esbjerg and Department for Thrombosis Research, Institute of Public Health, University of Southern Denmark, Esbjerg, Denmark; (2) EAA Central Facility, Faculty of Life Sciences, University of Manchester, Manchester, UK; (3) Department of Thrombosis and Haemostasis, Leiden University Medical Center, Leiden, The Netherlands; (4) Department of Thrombosis and Haemostasis, Leiden University Medical Center, Thrombosis Service Leiden, Leiden, The Netherlands; (5) Department of Angiology and Blood Coagulation, University Hospital Sant’Orsola-Malpighi, Bologna, Italy; (6) University and IRCCS Maggiore Hospital, Mangiagalli and Regina Elena Foundation, Milan, Italy; (7) Department of Haematology, Manchester Royal Infirmary, Oxford Road, Manchester, UK

Summary

Anticoagulant control facilities are being overwhelmed by requests for monitoring and large numbers of patients are not therefore receiving treatment. Procedures designed for point-of-care testing have therefore been developed, the most popular being the CoaguChek. The need for external quality assessment (EQA) of monitors used by patients in self-management has been stressed in a European Commission (EC) Directive. It would not however be feasible for all CoaguChek monitors to be enrolled in national or regional EQA schemes which take time to organise and analyse. The European Concerted Action on Anticoagulation (ECAA) has therefore evolved a simpler system. Its value has been assessed in collaboration with the European Concerted Action on Thrombosis (ECAT). 523 monitors were tested at nine clinics which asked patients to bring their CoaguChek instruments to be assessed with the ECAA/ECAT procedure based on a set of 5 plasma samples with certified international normalised ratios (INR). 15% or more deviation from the certified INR on a single certified plasma sample from the set was defined by the ECAA as the limit of acceptable performance. One hundred and six (20.3%) of the monitors tested showed significant deviation and higher than average incidence of significant INR deviations reported with one specific numbered lot of test strips. Recent ECAA/ECAT, Danish and Italian studies report regular EQA of CoaguChek monitors is essential. There is general agreement that this should be performed at reasonably frequent intervals, at six months or whenever there is a change of the manufacturer’s test strips.

Keywords

accuracy, precision, Oral anticoagultaion, external quality assessment, point-of-care monitors, INR deviation

DOI

http://dx.doi.org/10.1160/TH09-10-0683

You may also be interested in...

1.

M. Neuhäuser, T. Boes, K.-H. Jöckel

Methods of Information in Medicine 2007 46 5: 538-541

http://dx.doi.org/10.1160/ME0397

2.

P. Andermann, S. Schlögl, U. Mäder, M. Luster, M. Lassmann, Chr. Reiners

Nuklearmedizin 2007 46 1: 1-7

3.
Stand der Forschung und Verbesserungsoptionen

D. Kendoff (1, 2)

Die Medizinische Welt 2010 61 6: 338-346



Articles

You've 558 Article(s) in your Basket.

TH 107.5

Clinical Focus on GPIIb/IIIa inhibitors: In the May issue of Thrombosis and Haemostasis Armstrong...

TH 107.4

The April 2012 issue of Thrombosis and Haemostasis TH 107.4 is a Theme Issue by A. Schober, T....

Thrombosis and Haemostasis official organ of Spanish Society for Thrombosis and Haemostasis

Thrombosis and Haemostasis, founded in 1957, has become the official organ of the Spanish Society...