In vitro biomechanical comparison of limited contact dynamic compression plate and locking compression plate

Journal:Veterinary and Comparative Orthopaedics and Traumatology (VCOT)
ISSN:0932-0814
Issue:2005: Issue 4 2005
Pages:220-226

In vitro biomechanical comparison of limited contact dynamic compression plate and locking compression plate

A. Z. Aguila1 , J. M. Manos2 , A. S. Orlansky2 , R. J. Todhunter1 , E. J. Trotter 1 , M. C. H. van der Meulen2
1 College of Veterinary Medicine, Cornell University, Ithaca, New York, USA 2 Sibley School of Mechanical & Aerospace Engineering, Cornell University, Ithaca, New York, USA

Summary

The locking compression plate (LCP) supports biologicalosteosynthesis by functioning as an internal fixator,rather than as a full or limited contact bone plate whichmust be adequately contoured and affixed directly tothe bone for stable internal fixation of the fracture. Inorder to help justify the use of the LCP in our veterinarypatients, in vitro biomechanical testing was performedcomparing the LCP to the conventional limited contactdynamic compression plate (LC-DCP) in canine femurs.We hypothesized that the LCP construct would be atleast as stiff under bending and torsional loads as theLC-DCP. The LCP and LC-DCP were applied over a20-mm osteotomy gap to contralateral bones withineach pair of 14 femora. Non-destructive four-pointbending and torsion, and cyclical testing in torsion wereperformed. The constructs were then loaded to failure intorsion. In medial-lateral and lateral-medial structuralbending, significant differences were not found betweenthe LCP and LC-DCP, however, at the gap, the LCP constructwas stiffer than the LC-DCP in lateral-medialbending. Significant differences in behaviour over timewere not noted between the plate designs during cyclicaltesting. When loading the constructs to failure in internalrotation, the LC-DCP failed at a significantlylower twist angle (P = .0024) than the LCP. Based onthe similar performance with loading, the locking compressionplate is a good alternative implant for unstablediaphyseal femoral fracture repair in dogs.

Keywords

Mechanical testing, LCP, LC-DCP

You may also be interested in...

1.

M. DeTora1, K. Kraus2

VCOT - Veterinary and Comparative Orthopaedics and Traumatology 2008 21 4: 318-322

http://dx.doi.org/10.3415/VCOT-07-04-0034

2.

D. Filipowicz (1), O. Lanz (2), R. McLaughlin (3), S. Elder (4), S. Werre (4)

VCOT - Veterinary and Comparative Orthopaedics and Traumatology 2009 22: 270-277

http://dx.doi.org/10.3415/VCOT-08-05-0042

3.

K. Zahn1, R. Frei2, D. Wunderle2, B. Linke2, K. Schwieger2, B. Guerguiev2, O. Pohler3, U. Matis1

VCOT - Veterinary and Comparative Orthopaedics and Traumatology 2008 21 3: 185-194



Articles

You've 944 Article(s) in your Basket.

VCOT 3/2012

Synoviocentesis of the equine digital flexor tendon. In issue 3/2012 of VCOT, Jordana et al. give...

VCOT 2/2012

The need for revision in non-cemented total hip implants is a growing problem in people who are...

VCOT 1/2012

Cranial cruciate ligament rupture is one of the most common conditions affecting the stifle joint...