Biomechanical characteristics of allogeneic cortical bone pins designed for fracture fixation

Journal:Veterinary and Comparative Orthopaedics and Traumatology (VCOT)
ISSN:0932-0814
DOI:http://dx.doi.org/10.3415/VCOT-07-03-0026
Issue:2008: Issue 2 2008
Pages:140-146

Biomechanical characteristics of allogeneic cortical bone pins designed for fracture fixation

J. M. Liptak1, M. R. Edwards1, S. P. James2, W. S. Dernell1, R. J. Scott1, A. M. Bachand3, S. J. Withrow1
1Department of Clinical Sciences, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, Colorado, USA 2Department of Mechanical Engineering, College of Engineering, Colorado State University, Fort Collins, Colorado, USA 3Department of Environmental and Radiological Health Sciences, College of Veterinary Medicine and Biomedical Sciences, Colorado State University, Fort Collins, Colorado, USA

Summary

The biomechanical characteristics of 1.2 mm diameter allogeneic cortical bone pins harvested from the canine tibia were evaluated and compared to 1.1 mm diameter stainless steel pins and 1.3 mm diameter polydioxanone (PDS) pins using impact testing and four-point bending. The biomechanical performance of allogeneic cortical bone pins using impact testing was uniform with no significant differences between sites, side, and gender. In four-point bending, cortical bone pins harvested from the left tibia (204.8 ± 77.4 N/mm) were significantly stiffer than the right tibia (123.7 ± 54.4 N/mm, P=0.0001). The site of bone pin harvest also had a significant effect on stiffness, but this was dependent on interactions with gender and side. Site C in male dogs had the highest mean stiffness in the left tibia (224.4 ± 40.4 N/mm), but lowest stiffness in the right tibia (84.9 ± 24.2 N/mm). Site A in female dogs had the highest mean stiffness in the left tibia (344.9 ± 117.4 N/mm), but lowest stiffness in the right tibia (60.8 ± 3.7 N/mm). The raw and adjusted bending properties of 1.2 mm cortical bone pins were significantly better than 1.3 mm PDS pins, but significantly worse than 1.1 mm stainless steel pins (P<0.0001). In conclusion, cortical bone pins may be suitable as an implant for fracture fixation based on initial biomechanical comparison to stainless steel and PDS pins used in clinical practice.

Keywords

dog, biomechanics, Allogeneic, cortical bone, stainless steel, polydioxanone, pins

DOI

http://dx.doi.org/10.3415/VCOT-07-03-0026

You may also be interested in...

1.

A. Kaiser1, P. Böttcher2, J. Maierl3

Tierärztliche Praxis Kleintiere 2007 35 4: 261-266

2.

I. Jopp; S. Reese

VCOT - Veterinary and Comparative Orthopaedics and Traumatology 2009 22 2: 119-124

http://dx.doi.org/10.3415/VCOT-08-03-0029

3.

R. A. Pettitt, J. F. Innes

VCOT - Veterinary and Comparative Orthopaedics and Traumatology 2008 21 3: 302-306



Articles

You've 810 Article(s) in your Basket.

VCOT 3/2012

Synoviocentesis of the equine digital flexor tendon. In issue 3/2012 of VCOT, Jordana et al. give...

VCOT 2/2012

The need for revision in non-cemented total hip implants is a growing problem in people who are...

VCOT 1/2012

Cranial cruciate ligament rupture is one of the most common conditions affecting the stifle joint...