Evaluation of a short glass fibre-reinforced tube as a model for cat femur for biomechanical testing of orthopaedic implants

Journal:Veterinary and Comparative Orthopaedics and Traumatology (VCOT)
ISSN:0932-0814
Issue:2008: Issue 3 2008
Pages:195-201

Evaluation of a short glass fibre-reinforced tube as a model for cat femur for biomechanical testing of orthopaedic implants

T. W. G. Gibson1, N. M. M. Moens1, R. J. Runciman2, D. L. Holmberg1

1Ontario Veterinary College, Department of Clinical Studies; 2School of Engineering, University of Guelph, Guelph, Ontario, Canada

Summary

The biomechanical testing of tubes made of third generation short glass fibre-reinforced (SGFR) material approximating cat femurs was performed in order to determine their suitability as cat femur surrogates for the biomechanical testing of orthopaedic implants. The tubes were tested in compression, three-point bending, notch testing, and screw pullout. Thin walled (B1-tubes) had a 13% lower maximum load to failure, a 19% higher maximum strength and a 13% lower elastic modulus compared to cat femurs tested in compression. B1-tubes maximum load to failure in threepoint bending and screw pullout strength were considerably lower compared to cat femurs (29% and 63%, respectively). Notch testing was not performed on B1-tubes due to low bending strength. Thicker walled (B2-tubes) had a 23% higher maximum load to failure, a 10% higher maximum strength and a 21% lower elastic modulus compared to cat femurs tested in compression. The comparison of B2-tubes and cat femurs in three-point bending revealed a 7% increase in maximum load to failure for the B2-tubes. Drilled B2-tubes (notch testing) were weaker with a 30% lower load to failure compared to cat femurs. A screw pullout comparison of B2-tubes and cat femurs revealed a 2% increase in maximum load to failure for the B2-tubes. These tubes were intended to provide a model as a suitable surrogate for cat femurs for testing the bending strength of various orthopaedic constructs involving plates and screws. Testing revealed that third generation SGFR tubes were not suitable for these purposes and emphasizes the need to carefully evaluate the suitability of any model.

Keywords

compression, Cat femur, three-point bending, Short glass fibre-reinforced material, biomechanical testing, notch testing

You may also be interested in...

1.

T. W. G. Gibson1, N. M. M. Moens1, R. J. Runciman2, D. L. Holmberg†, G. M. Monteith1

VCOT - Veterinary and Comparative Orthopaedics and Traumatology 2008 21 4: 312-317

http://dx.doi.org/10.3415/VCOT-07-08-0077

2.

Y. Herouy (1), B. Kahle (2), M. Idzko (3), J. Norgauer (4), E. Rabe (5), L. Bruckner-Tuderman (6), M. Jünger (7)

Phlebologie 2004 33 4: 115-119

3.

R. Bisacci1, P. De Rango2, C. Bisacci1, V. Pagliuca1, G. M. Spinelli1

Phlebologie 2007 36 6: 293-302



Articles

You've 950 Article(s) in your Basket.

VCOT 3/2012

Synoviocentesis of the equine digital flexor tendon. In issue 3/2012 of VCOT, Jordana et al. give...

VCOT 2/2012

The need for revision in non-cemented total hip implants is a growing problem in people who are...

VCOT 1/2012

Cranial cruciate ligament rupture is one of the most common conditions affecting the stifle joint...