In vitro biomechanical testing of a micro external skeletal fixator

Journal: Veterinary and Comparative Orthopaedics and Traumatology (VCOT)
ISSN: 0932-0814
DOI: http://dx.doi.org/10.3415/VCOT-12-10-0126
Issue: 2013: Issue 5 2013
Pages: 385-391

In vitro biomechanical testing of a micro external skeletal fixator

Online Supplementary Material

R. Deiss (1), M. S. Bali (2), M. Doherr (3), D. Spreng (4), U. Rytz (4), S. J. Ferguson (5)

(1) Veterinary Practice Kiesen, Kiesen, Switzerland; (2) Small Animal Hospital Stommeln, Stommeln, Germany; (3) University of Bern, Veterinary Medicine, Clinical Research, Bern, Switzerland; (4) University of Bern, Small Animal Clinic, Bern, Switzerland; (5) Swiss Federal Institute of Technology Zurich (ETH), Institute for Biomechanics, Zurich, Switzerland

Keywords

cats, External skeletal fixation, miniature dogs, frame configuration, close pin placement

Summary

Objective: To biomechanically test the properties of three different Universal Micro External Fixator (UMEX™) configurations with regard to their use in very small animals (<5kg) and compare the UMEX system to the widely used IMEX External Skeletal Fixation (SK™) system in terms of stiffness, space needed for pin placement and weight. Methods: Three different UMEX configurations (type Ia, type Ib, and type II modified) and one SK configuration type Ia were used to stabilize Delrin plastic rods in a 1 cm fracture gap model. These constructs were tested in axial compression, craniocaudal bending, mediolateral bending, and torsion. Testing was conducted within the elastic range and mean stiffness in each mode was determined from the slope of the linear portion of the load-deformation curve. A Kruskal Wallis one-way analysis of variance on ranks test was utilized to assess differences between constructs (p <0.05). Results: The UMEX type II modified configuration was significantly stiffer than the other UMEX configurations and the SK type Ia, except in craniocaudal bending, where the SK type Ia configuration was stiffer than all UMEX constructs. The UMEX type Ia configuration was significantly the weakest of those frames. The UMEX constructs were lighter and smaller than the SK, thus facilitating closer pin placement. Conclusions: Results supported previous reports concerning the superiority of more complex constructs regarding stiffness. The UMEX system appears to be a valid alternative for the treatment of long-bone fractures in very small animals.

You may also be interested in...

1.
F. Meige1, S. Sarrau2, A. Autefage3

Vet Comp Orthop Traumatol 2008 21 1: 76-84

http://dx.doi.org/10.1160/VCOT-07-01-0010

2.
S. Lauer, G. Hosgood, S. Ramirez, M. Lopez

Vet Comp Orthop Traumatol 2008 21 1: 25-35

http://dx.doi.org/10.3415/VCOT-06-11-0090

3.
H. F. Reaugh1, M. C. Rochat2, C. W. Bruce3, D. S. Galloway4, M. E. Payton5

Vet Comp Orthop Traumatol 2007 20 4: 264-268

http://dx.doi.org/10.1160/VCOT-06-05-0045