The use of the VerifyNow P2Y12 point-of-care device to monitor platelet function across a range of P2Y12 inhibition levels following prasugrel and clopidogrel administration

Journal:Thrombosis and Haemostasis
ISSN:0340-6245
DOI:http://dx.doi.org/10.1160/TH07-09-0575
Issue:2008: 99/2 (Feb) pp. 249-455
Pages:409-415

The use of the VerifyNow P2Y12 point-of-care device to monitor platelet function across a range of P2Y12 inhibition levels following prasugrel and clopidogrel administration

Joseph A. Jakubowski1, Christopher D. Payne2, Ying G. Li1, John T. Brandt1, David S. Small1, Nagy A. Farid1, Daniel E. Salazar3, Kenneth J. Winters1
1Lilly Research Laboratories, Eli Lilly and Company, Indianapolis, Indiana, USA; 2Lilly Research Laboratories, Eli Lilly and Company, Windlesham, Surrey, UK; 3Daiichi Sankyo, Inc. Parsippany, New Jersey, USA

Summary

Variability in response to antiplatelet agents has prompted the development of point-of-care (POC) technology.In this study,we compared theVerifyNow P2Y12 (VN-P2Y12) POC device with light transmission aggregometry (LTA) in subjects switched directly from clopidogrel to prasugrel. Healthy subjects on aspirin were administered a clopidogrel 600 mg loading dose (LD) followed by a 75 mg/d maintenance dose (MD) for 10 days. Subjects were then switched to a prasugrel 60 mg LD and then 10 mg/d MD for 10 days (n=16), or to a prasugrel 10 mg/d MD for 11 days (n=19). Platelet function was measured by LTA andVN-P2Y12 at baseline and after dosing. Clopidogrel 600 mg LD/75 mg MD treatment led to a reduction in P2Y12 reaction units (PRU) from baseline. A switch from clopidogrel MD to prasugrel 60 mg LD/10 mg MD produced an immediate decrease in PRU, while a switch to prasugrel 10 mg MD resulted in a more gradual decline. Consistent with the reduction in PRU, device-reported percent inhibition increased during both clopidogrel and prasugrel regimens. Inhibition of platelet aggregation as measured by LTA showed a very similar pattern to that found with VN-P2Y12 measurement, irrespective of treatment regimens.The dynamic range of VN-P2Y12 appeared to be narrower than that of LTA. With two different thienopyridines, the VN-P2Y12 device, within a somewhat more limited range, reflected the overall magnitude of change in aggregation response determined by LTA.The determination of the clinical utility of such POC devices will require their use in clinical outcome studies.

Keywords

clopidogrel, platelet aggregation, prasugrel, VerifyNow™ P2Y12, light transmission aggregometry

DOI

http://dx.doi.org/10.1160/TH07-09-0575

You may also be interested in...

1.

Pancras C. Wong; Earl J. Crain; Carol A. Watson; Ji Hua; William A. Schumacher; Robert Rehfuss

Thrombosis and Haemostasis 2009 101 1: 108-115

http://dx.doi.org/10.1160/TH08-07-0456

2.

Christian Gachet

Thrombosis and Haemostasis 2008 99 3: 466-472

http://dx.doi.org/10.1160/TH07-11-0673

3.

K. Schrör1, K. Huber2

Hämostaseologie 2007 27 5: 351-355



Articles

You've 1004 Article(s) in your Basket.

TH 107.5

Clinical Focus on GPIIb/IIIa inhibitors: In the May issue of Thrombosis and Haemostasis Armstrong...

TH 107.4

The April 2012 issue of Thrombosis and Haemostasis TH 107.4 is a Theme Issue by A. Schober, T....

Thrombosis and Haemostasis official organ of Spanish Society for Thrombosis and Haemostasis

Thrombosis and Haemostasis, founded in 1957, has become the official organ of the Spanish Society...